Infusion Pump
Available on request
- SKU
- 63
Features of Infusion Pump
- HD LCD Display, high capacity words, friendly user interface, dynamically display working status;
- Audible and visual alarm for occlusion, empty, low battery, end of infusion, door open, wrong setting, etc., which gain patents;
- Compatible with any brands of infusion sets after correct calibration;
- Preset solution volume to greatly reduce the workload of nurses;
- Work mode: ml/h and drop/min can switch freely;
- Three levels of occlusion: high, middle and low;
- Purge function;
- KVO (keep-vein-open) automatically opens as infusion is completed, KVO rate is 1-5ml/h(1ml/h step);
- Power Source: AC100---240V, 50/60Hz; Internal Battery;
- Automatically record the settings of last infusion;
Specifications
|
Rate |
1ml/h ~ 1,200ml/h |
|
Flow Rate Accuracy |
Within ±5% (after correct calibration) |
|
Mechanical precision |
Within ±2% |
|
Purge Rate |
100ml/h ~ 1,000ml/h (100 ml/h step) |
|
Infusion Volume |
1ml ~ 9999ml |
|
Total Infusion Volume |
0.1ml ~ 9999.9ml |
|
KVO rate |
1ml/h ~ 5 ml/h (1ml/h step) |
|
Occlusion |
High: 800mmHg ±200mmHg (106.7kPa±26.7kPa |
|
Unique audible and visible alarm |
Human voice alarm for End of injection, occlusion, door open, bubbles in the tube, wrong setting, low battery, AC power pulled out etc. |
|
Power Source |
AC 100V ~ 240V, 50/60Hz; Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup |
|
Bubble detector |
Ultrasonic wave detector; detection sensitivity ≥25μL |
|
Fuse |
F1AL/250V, 2 pc inside |
|
Power Consumption |
18VA |
|
Operating Condition |
Ambient temperature: +5℃ ~ +40℃; |
|
Transport & Storage Condition |
Ambient temperature:-30℃ ~ +55℃ |
|
Equipment Classification |
Class II, internal power supply, Type CF |
|
IP Classification |
IPX4 |
|
Dimension |
153mm (L) ×162mm (W) × 227mm (H) |
|
Weight |
1.8kg (net weight) |